Closed ciozi137 closed 2 years ago

ciozi137 commented 2 years ago

2 days in a row on PPMS3 DAQ: (memory leak?) v1.8.8.8

image

ciozi137 commented 2 years ago

image

ciozi137 commented 2 years ago

Memory leak for sure. Errors are coming from Remote Control process: image

ciozi137 commented 2 years ago

image

ciozi137 commented 2 years ago
Date    Time    Type    Name    Code    Message Possible reason(s)  Call Chain  Calling Class
2022-09-11  19:11:41.797    Err     13  Error 13 occurred at Call Library Function Node in zeromq.lvlib:zmq_context.lvclass:zmq_ctx_create.vi:5610003->RemoteControl.ZMQ.lvclass:Open Server Connection.vi->RemoteControl.lvclass:Process.vi:5310001->SMO.lvclass:LaunchProcess.vi:3770001->SMO.lvclass:LaunchProcess.vi.ACBRProxyCaller.91B0002E   LabVIEW: (Hex 0xD) Failed to load dynamic library because of missing external symbols or dependencies, or because of an invalid file format.
=========================

Fail to load dll...problem with build...build spec?

ciozi137 commented 2 years ago

Turns out it was missing or corrupt 2013 C++ dll