Make sure lockin freq. is 1Hz ( control expt. create output malfunctions if the lockin is updateing slowly)
Set up control expt and press create output wait for KH to create output
Put a given input e.g 20mV at I1+,I2+,I3+
Reset Lockin freq. to 20mHz, also adjust the sampling rate accordingly
Press zero offset/measure offset
• Make sure Lockin is not in idle mode?
Go to “sweep control” set the x gate values (triangle wavefunctions)
• Priority 0
• Amplification factor?
• Start end
• Step size
• channel
• Wait? wait for few seconds after setting a gate value for system to stabilize
• Post action: do nothing
Set number of IV at each gate
Run sweep control for taking IV measurements
Repeat for Y gate and then XY gate
• Same set up as X?
• XY requires 4 channels instead of 2
• Channel number changes
• Expt. Name changes
• Comments changes (automate comments? Standardize comments from control expt?)
Repeat steps 1-9 for the remaining three sides of the three devices
Repeat steps 1-10 at a different magnetic field (for first set can avoid step 1-4?)
• Set start and end and step size
• Post action: Check magnet field value and temperature
• Magnet in persistent or driven mode?
Change magnetic field
• Repeat steps 1-11
Put Magnet in persistent mode at last field value?