Closed 0xNoor closed 1 year ago
After reinstalling the app, and not using auto fixed the issue. The main cause of the issue is the voltage/division. As you can see by using probe at x10, the V/div is ±24 V and the ADC resolution is 8bit of OWON VDS1022 which means the scope would measure every interval of 187.5 mV at ±24 V . Since my Vpp is around 970mV in sine wave, so I got around ~4 readings.
I am using OWON VDS1022 and realme 8 together to use HScope. On using the probe at x10, the results are apearing to be very jagged. I tired confirming if this is a hardware or software problem, and confirmed that this is surely a problem with Hscope.
At x1 probe florentbr/OWON-VDS1022 software - 440hz Triagular wave
At x10 probe florentbr/OWON-VDS1022 software - 440hz Triagular wave
Hscope x1 probe - 440hz Triagular wave
Hscope X10 probe - 440hz Triagular wave
Hscope x10 probe - 1Khz sine wave
Sine waves are looking pretty awful at x10 probe