adjusted max resist alarm and AC and BPM knobs #62

Closed abukhalaf closed 4 years ago

abukhalaf commented 4 years ago

Lowering BPM min to 6 to better allow for assist control to sync with a patient. Increases the range of AC from 2 to 7, instead of 2 to 5. Also do this at increments of 1. Increase threshold of resist pressure alarm to 10cm H2O to reduce false alarms.

braraki commented 4 years ago

Can the increments for AC trigger be 0.5 instead of 1? When Albert and I talked last week, that's what we decided on. If you've talked with him since then and he said 1, then that would take precedence.

abukhalaf commented 4 years ago

@braraki AC increments changed to 0.5.

abukhalaf commented 4 years ago

@amadoantonini In the documentation at https://e-vent.mit.edu/clinical/key-ventilation-specifications/ , we have added earlier this "Respiratory Rate (RR) (breaths per minute): between 6 – 40. Note that the low RRs of 6 – 9 are only applicable to Assist Control."

amadoantonini commented 4 years ago

Ok, thanks for the clarification.