After receiving feedback from clinicians, we have settled on a new display format that will allow us to incorporate more information in a more compact layout.
See below for images showing the new functionality. The key feedback we received includes:
Show the status/alarm on the top (A) and the icons / setting change on the bottom (C)
The icons for Pip/Plat/Peep using waveforms were understood and preferred to arrows
the 1:x should be shown for IE ratio, and also included in the I:E (decided to compress it into a single glyph for space)
I:E ratio should be allowed to go as low as 1:0.4
The time-triggered vs patient-triggered icons were understood, and should be placed above the pressures (showin below in top-right corner)
AC knob should be labeled "Ptr"
Alarm sound icon should only show when an alarm is sounding (B), otherwise, it should be blank (A)
When an alarm is occurring but muted, the sound icon should indicate this, and also count down seconds until muting will stop (C).
After receiving feedback from clinicians, we have settled on a new display format that will allow us to incorporate more information in a more compact layout.
See below for images showing the new functionality. The key feedback we received includes:
A: Normal Running
B: Alarming
C: Change Settings