Interlock message is not responding appropriately #69

Open lo-co opened 8 years ago

lo-co commented 8 years ago

Always says lasers are contained.

JustinLangridge commented 8 years ago

In addition part of lthe laser safety case is that the lasers will be powered 1 by 1 when the lid is off. INd addition to updating the interlock message, can the software set either all laser powers to 0% or turn off all laser enable signals (if we get them working) when it senses the interlock has been triggered

lo-co commented 8 years ago

Hmmmm...is this tested in MAX? I will take a look at.

The Interlock is monitored in the Controller. We will send a message if we see the interlock is active to shutdown the lasers.

lo-co commented 8 years ago

Updated sidebar controller so now it indicates what is actually going on with the interconnect. Now need to provide proper indication of when a laser is on. Laser is on when: