msr-consulting / exscalabar

User interface for the EXSCALABAR instrument.
http://www.msrconsults.com/ukmet-gh/exscalabar
1 stars 2 forks source link

UI control initialistation problem #89

Closed JustinLangridge closed 8 years ago

JustinLangridge commented 8 years ago

While testing the CRD laser digital pulse signal, it appears that the UI controls for the frequency are not initializing correctly.

E.g. on launching the UI the red lasers frequency is set to 1kHz. This value matches that shown in the exscalabar.ini file. However the actual frequency of the pulse coming out the PXI is 1.3kHz. When the user manually re-enters 1kHz in the control, the output waveform then changes to the 1kHz value.

A similar issue is seen for the blue laser control (and maybe is a problem for other UI controls?)

Why could this be and can we fix it?

lo-co commented 8 years ago

Is this the same for both lasers, or just the red? This is likely associated with lo-co/exscalabar_server#124. Closing this as it is not an interface issue.