nightscout / AndroidAPS

Opensource automated insulin delivery system (closed loop)
https://wiki.aaps.app
GNU Affero General Public License v3.0
696 stars 1.7k forks source link

TBR set for 120 seconds with OrangeLink instead of 120 minutes #1887

Closed Muppet1313 closed 2 years ago

Muppet1313 commented 2 years ago

Hardware / Software ;

Samsung A52 5G AAPS version 3.0.0.2 OrangeLink hardware Version 1.0 with Firmware v3.2 Medtronic Veo 554

Reporting bugs

Pump sets 120 minutes Temporary Basal rate 0.000 U for 120 minutes-> 08:14:52.614 [Thread-15922] I/PUMPCOMM: [MedtronicCommunicationManager.setTemporaryBasal():611]: setTBR: Rate: 0.000 U, Duration: 120 min 08:14:54.125 [Thread-15922] I/PUMP: [MedtronicPumpPlugin.setTempBasalAbsolute():761]: MedtronicPumpPlugin::setTempBasalAbsolute - setTBR. Response: true

NS says 120 seconds ->

08:14:54.134 [Thread-15922] D/CORE: [PumpSyncStorage.addTemporaryBasalRateWithTempId():179]: PumpDbEntryTBR: PumpDbEntryTBR(temporaryId=20220705081454, date=1657001694125, pumpType=MEDTRONIC_554_754_VEO, serialNumber=856911, pumpId=null, rate=0.0, isAbsolute=true, durationInSeconds=120, tbrType=NORMAL)

AndroidAPS 3 0 0 2 AndroidAPS_LOG_1657001820157.log.zip screenshot 4-7-22 NS - no basal stop

andyrozman commented 2 years ago

Are you sure this doesn't work? I haven't made any changes in the code there.... So when this is set:

  1. What is set on pump?
  2. What is shown in Treatments (in AAPS)
  3. And what is shown in NS...
Muppet1313 commented 2 years ago

unfortunately I can't reproduce it now the error message because we have switched to the Dana I pump. I will close the ticket.