Manual AAPS delivered bolus not picked up in treatments, may not be in IOB #253

Closed tim2000s closed 2 years ago

tim2000s commented 3 years ago

Reporting bugs

21.05 - manual bolus administered via AAPS and showing in pump history not showing in treatments or uploaded to Nightscout. Not clear that it's included in IOB. IOB may be undercounted.

At the time AAPS issued a basal not updated followed by a bolus not delivered alert.

Pump in use is Medtronic 722. Logs uploaded. /storage/emulated/0/Android/data/info.nightscout.androidaps/ See https://androidaps.readthedocs.io/en/latest/EN/Usage/Accessing-logfiles.html

tim2000s commented 3 years ago

Ran OpenAPS rig against the pump and confirmed that the bolus was missing from IOB.

andyrozman commented 3 years ago

That is weird... When you administer bolus from AAPS this information is added to treatments, unless something went wrong... Even if it wasn't added to treatment, if it was delivered with pump, it should have been picked up from history...

Can you please provide logs @tim2000s

MilosKozak commented 3 years ago

we don't have logs

tim2000s commented 3 years ago

That's odd. They were sent through last night. Please let me know if you haven't received them by now.

tim2000s commented 3 years ago

I've resent the logs through just now. To the usual logs@androidaps.org address.

AdrianLxM commented 3 years ago

@tim2000s androidaps.org email addresses do no longer work.

tim2000s commented 3 years ago

@tim2000s androidaps.org email addresses do no longer work.

That's the address that AAPS used from within the system...

Files uploaded below. Original files were wrong date.

AndroidAPS._2021-01-2200-00-13.24.zip AndroidAPS._2021-01-2200-00-13.25.zip AndroidAPS._2021-01-2200-00-13.26.zip

StephenBrown2 commented 3 years ago

What email should be used? I have sent several logs via the Maintenance menu, now I think they must have gone nowhere, though I haven't gotten any bounce responses.

andyrozman commented 2 years ago

@tim2000s When dev is released, please retest this... I have reworked the whole Medtronic module (converted into kotilin), together with database integration changes. There were also some changes with history data handling, so this might have been fixed in the meantime.