Open jonnew opened 4 days ago
@bparks13
It seems like there is a little left over icon at the bottom of the GUI still:
When i run the GUI on a ConfigureNeuropixelsV1f
(no headstage), there is a "Probe A" indicator in the window title:
I dont think that should be there because the A vs. B is given by the device address in this case. Basically, nothing needs to change compared to the ConfigureNeuropixelsV1e
in this case.
@jonnew The icon is supposed to be there, it indicates if the calibration files are valid, and if the serial numbers of the files match. This does not touch hardware, but is meant as an easy way to see which files are currently loaded and if they match.
However, the text is being cut off, which is incorrect, so that needs to be fixed.
When I see the circle arrow icon, I expect to be able to click it to refresh, but that does not seem to be the case here. I would be a yellow warning triagle or something to indicate that cal files are needed