Rx Dose for XiO plans #78

Open GoogleCodeExporter opened 9 years ago

GoogleCodeExporter commented 9 years ago
The absolute dose matrix in XiO RTDOSE files can be calculated by taking the 
pixel data of the dose cube (which typically has a range approximately 0-1e6 
corresponding to 0-100%) and multiplying it by the RTDOSE's DICOM field 
(3004,000E) Dose Grid Scaling as described in 
ftp://medical.nema.org/medical/dicom/final/cp221_ft.pdf. Would it be possible 
to implement this? Perhaps some of the other software that currently requires 
Rx Dose to be input manually also uses the same field for scaling.

Thanks for making a very clean and useful piece of software!

Dan

Original issue reported on code.google.com by DanielRWarren25 on 19 Jul 2012 at 4:16

GoogleCodeExporter commented 9 years ago
The next version will support Rx Dose that is derived from the fractionation 
information within an RT Plan file. So far, this has covered all of the cases 
that I have encountered where the Rx Dose is not present.

I think your suggestion is a good one to fall back when an RT Plan is either 
not available or does not contain fractionation information.

Thanks for the kind comments about dicompyler Dan!

Adit