Closed 4psireal2 closed 1 year ago
QCxMS works with two general modes of operation, one for EI-MS (GC-MS) and one for CID-MS/MS (LC-MS/MS). Depending on that you need to choose your input molecule and parameter set.
@psireal9 Your example looks like N-potonated codeine, as [M+H]+ adduct with the molecular formula C18H21NO3+ Meaning you want to calculate the CID-MS/MS? Then you should try with the standard CID-MS/MS settings:
xtb2
cid
elab 60
maxcoll 6
noesi
In case you want to calculate the 70eV EI-MS (for GC-MS) spectrum, you need to remove the proton and add the neutral molecule as optimized 3D.
I ran
qcxms
for the first time and by the second runqcxms
, I got this errorThe
coord
andqcxms.in
input files: coord.txt qcxms.in.txt