Closed nkrackow closed 9 months ago
Ah I see we have the data more or less in https://github.com/quartiq/thermostat-eem/issues/38#issuecomment-1127683289 with slightly different conditions. But that should not matter. Summary is that there is negligible benefit of the balanced measurement in this case. The next step would be A firmware version that only supports 16 single ended sensors (i.e. 5 wire measurement with 2 excitation, and 2 reference per adc plus one signal sense wire per sensor).
@nkrackow done now?
done
It would be really interesting to see the raw data for the single ended setup. For that with the current afe just change the ADC setup to 4 single ended channels everywhere (plus the impact on sequencing, data arrays, and timing), short the NTC and measure 5k ref vs the other 5k ref ratiometric.