Closed jordens closed 5 years ago
Agreed. Floating grounds can be the right thing to do in rare cases, but it's not a sensible default option.
(I thought we agreed to do this ages ago)
We agreed. I'm just pushing to rush this into the current boards without waiting for v1.2/v2.0.
a87c75c
Thanks. The v1.1.1 version you assigned is a very good idea.
I have seen a couple times that that floating shield thing is a bad idea, also on crate-internal MMCX wiring. The symptom is 400/800 kHz switcher modulation sidebands anywhere from -30 dBc to -80dBc on the urukul clock. Shorting the shield on both ends resolved it each time. This can a pure BOM change and I'd like to have it implemented on all future Kasli v1.1 builds and Kasli v2.0 as well.
R29,R30,R31,R32,R214=0R
c.f. #5 and maybe #10