Closed ghost closed 4 years ago
We found that a 24V supply was inadequate to drive more than ~4.5 amps in the low field scanner gradient coils. We added a second 24V in series. Buffer file is updated, and some capacitors and resistors need to be changed for the new ~50V VCC
Requires new buffer design design (issue linked)
Buffer is done, waiting on components and current-sense wiring
Waiting on capacitors to replace (from finished #44)
Ready for an image
We found that a 24V supply was inadequate to drive more than ~4.5 amps in the low field scanner gradient coils. We added a second 24V in series. Buffer file is updated, and some capacitors and resistors need to be changed for the new ~50V VCC