Here below the relevant QCs for the efficiency group
efficiency:
Relative efficiency for UVB arm, range 5100A - 5300
EFFMON template .. was observation started under photometric conditions?
median star counts in arm at given range
Chromatic efficiency plot for UVB arm; the values trended in plot are marked. The plot is constituted from the following wavelength ranges: 5670A, a point each circa 200 angstroms (chromatic efficiency)
-flat_response:
For each calibration flat lamp: D2 current response / reference response = a + b * (lam - lam0)
On 2015-10-25, the first flux standard star was acquired with the effmon template, meaning this observation was acquired under strict photometric conditions. The calibration product 'response function' from that date serves as a reference in the following. It is marked by a vertical red bar in the history plot.
The master response function once generated as a weighted and clean mean of response functions for the purpose to flux calibrate XSHOOTER science observations (internal data producs = IDPs) is a representative response function for a given time range. The comparison between the reference response function (from 2015-10-25, ARCFILE=XSHOO.2015-10-26T01:23...) and the current response function as well as the comparison between the master response and the current response is parameterized via
current(lambda) / reference(lambda) = a + b * (lambda - lambda_0 )
current(lambda) / master(lambda) = a + b * ( lambda - lambda_0 )
where lambda_0 is the center wavelength of the considered wavelength range, parameter a is the relative overall flux of the spectrum, corresponding to atmospheric variations and parameter b is a first order chromatic deviation (the slope between the reference(/master) and the current).
This plot shows both parameters a and b, of which parameter b is the important one.
On 2015-10-25, there is a=1 and b=0 by definition. The response function is a flat-fielded efficiency, hence the response function depends on the spectral shape of the flat field lamp, on long-term changes as well as on exchanges of the lamp. For this reason, dates of lamp exchanges are marked by blue vertical bars.
Here below the relevant QCs for the efficiency group
-flat_response: