Description
When PEEP setting(in the display, not the mechanical valve) is set to 1 and TV is set to 400, Tvi and Tve values fluctuated between 430-480
When PEEP setting(in the display, not the mechanical valve) is set to 10 and TV is set to 400, Tvi and Tve values fluctuated between 500-540
Description When PEEP setting(in the display, not the mechanical valve) is set to 1 and TV is set to 400, Tvi and Tve values fluctuated between 430-480 When PEEP setting(in the display, not the mechanical valve) is set to 10 and TV is set to 400, Tvi and Tve values fluctuated between 500-540
These co-relations should not exist.
Observations https://youtu.be/-dIWVRNNjQg https://youtu.be/YJWb0lTcY0o
Steps to Reproduce Set mechinical PEEP valve to 10 Set PEEP on display to 1 Set TV to 400 Observe Tvi and Tve
Set mechinical PEEP valve to 10 Set PEEP on display to 9 Set TV to 400 Observe Tvi and Tve