Open sudo-goatboy opened 4 years ago
Description- Home position, sensed by the inductive sensor skips ahead after 15-20 minutes of home position reset.
Observations- Regular home position : https://youtu.be/EiJBtRNucDE Skipped home position: https://youtu.be/vnJEr2SlJrs
Steps to Reproduce- Start machine Start ventilation Wait for 15-30 mins
Code Base- issue38-peep
This is fixed by adding extra 5 mm travel in exhale cycle in MCU1 code (relay module), this will make sure it travels extra and home is sensed every time, we have tested this for 15hour
Description- Home position, sensed by the inductive sensor skips ahead after 15-20 minutes of home position reset.
Observations- Regular home position : https://youtu.be/EiJBtRNucDE Skipped home position: https://youtu.be/vnJEr2SlJrs
Steps to Reproduce- Start machine Start ventilation Wait for 15-30 mins
Code Base- issue38-peep