Open NicoDrou opened 8 months ago
Hi Nicolas,
Intensities are reported in 'arbitrary units' for relative quantification, under 1000 is fine. The protein intensities are reported normalised by default, hence being similar for a dilution series is a good thing :)
Best, Vadim
OMG, indeed, I didn't notice the non-normalized intensities were in accordance with the dilution factor. IS there the same "arbitrary units" effect for other vendors or is it specific to Sciex? I am also using DIA-NN on TIMS-TOF data and i never had this feeling. On the contrary, protein intensities spawn over multiple orders of magnitude.
All vendors have different intensity scales.
Dear Vadim,
I have some issue to treat data acquired with a Sciex Zeno-SWATH. I did copy all the MS-convert files containing Sciex or core in their name into the DIA-NN folder. I also converted the data into .dia. The search was successful, with all the reports being generated and thousands of proteins detected as expected. However, the protein intensities cannot be correct. Except a few, all intensities are under 1000 and for a dilution series from 1 to 500ng, the detected proteins all have the same intensities, But trust me, the TIC followed the dilution trend :)
Did I forgot to add a command line?
Please find enclose the log file as well as the unique gene report. Zeno-swath.zip
Many thanks in advance for your help
Regards, Nicolas