Open patricktokeeffe opened 4 years ago
I fixed this issue by changing the analog integration time to 5000. Also, for Group A, the analog integration has to be at least 15000 to not report NaN.
Just following up here.. did those changes resolve the issue long-term? After revisiting, this looks like a bus address conflict or error since C4 data winds up in C3 but everything else gets a NAN.
Group C has 3 types of LICOR Biosciences open-path infrared gas analyzers:
The C3 irga is not installed, but every several dozen records is getting populated with data anyway. At the same time that occurs other SDM devices get bad data: NAN (LI7500 and CSAT3B) or -99999 (LI7500A).
Worth noting the LI7500 does not support group trigger,
CS7500()
trigger option 7