PROTO-TYPE-SOLUTIONS / easymed

Repository for Make-Easy HMIS. The all-in-one HMIS for Healthcare facilities.
12 stars 7 forks source link

LIS Integration for HumaCount 5D and HumaLyte Plus 3 #477

Open mosesmbadi opened 8 months ago

mosesmbadi commented 8 months ago

LIS Integration for:

HumaCount 5D HC5D_LISInterfaceManual_V1.pdf

HumaLyte Plus 3

Humalyte-Plus3-LIS-InterfaceManual_V1.pdf

Maglumi 600

HumaStar 100

arokiaswamy commented 5 months ago

Hi @mosesmbadi ,

Do you want to integration this machine with your LIS

mosesmbadi commented 2 months ago

Hi @mosesmbadi ,

Do you want to integration this machine with your LIS

Yes

dayaDanya commented 1 month ago

LIS Integration for:

HumaCount 5D HC5D_LISInterfaceManual_V1.pdf

HumaLyte Plus 3

Humalyte-Plus3-LIS-InterfaceManual_V1.pdf

Maglumi 600

HumaStar 100

Hello! Now I'm trying to create LIS Integration with HumaCount 80ts analyzer. I've already implemented messages receiving, but I have problems with sending hl7 messages to analyzer, it does not reply. May be you have any documentation or advice to solve this problem? I would appreciate it.

arokiaswamy commented 1 month ago

Hi,

I have the integration solution. If you want, I will do it for you. Each machine will charge $1000.

I have 15 years of experience in this field.

Thanks Arokia Swamy.

Regards,

Arokia Swamy S

On Wed, 25 Sep 2024 at 10:13 PM Moses Mbadi @.***> wrote:

Hi @mosesmbadi https://github.com/mosesmbadi ,

Do you want to integration this machine with your LIS

Yes

— Reply to this email directly, view it on GitHub https://github.com/PROTO-TYPE-SOLUTIONS/easymed/issues/477#issuecomment-2374836006, or unsubscribe https://github.com/notifications/unsubscribe-auth/ACCRDLQLLYCWDK2BT2FUZODZYL4OPAVCNFSM6AAAAABJBYQCIGVHI2DSMVQWIX3LMV43OSLTON2WKQ3PNVWWK3TUHMZDGNZUHAZTMMBQGY . You are receiving this because you commented.Message ID: @.***>

mosesmbadi commented 1 month ago

LIS Integration for: HumaCount 5D HC5D_LISInterfaceManual_V1.pdf HumaLyte Plus 3 Humalyte-Plus3-LIS-InterfaceManual_V1.pdf Maglumi 600 HumaStar 100

Hello! Now I'm trying to create LIS Integration with HumaCount 80ts analyzer. I've already implemented messages receiving, but I have problems with sending hl7 messages to analyzer, it does not reply. May be you have any documentation or advice to solve this problem? I would appreciate it.

@dayaDanya The HL7 message you are sending to the equipment should be correctly formatted. First, try to establish that the analyzer is communication with your LIS before sending the data. Then send the data and make sure it's formatted as explained in the guide attached. You can share your code I take a look.

arokiaswamy commented 1 month ago

Hi Moses,

HumaLyte Plus - This machine does not support bi-directional based on the manual what you attached.

Try the below response message once you receive the query message. Edit the message as per your

HumaCount 5D - Query Response message

MSH|^~\&|LIS||||20081120174836||ORR^O02|1|P|2.3.1||||||UNICODE MSA|AA|"MsgId" // Should be from ORM message PID|1||SAMPLEID^^^^MR||^" + PatientName + "||" + PatientDOB + "|" + PatientSex PV1|1|nk^^Bn4|||||||||||||||||Self-paid ORC|AF|SAMPLEID||| OBR|1|SAMPLEID||||20060506||||tester|||Notes....|20060504||||||||20080821||HM||||Validator||||Operator // Try this if the below like messae not working - OBX|1|IS|08001^Take Mode^99MRC||A OBX|1|IS|08001^Loading Mode^99MRC||A||||||F OBX|2|IS|08002^Blood Mode^99MRC||W||||||F OBX|3|IS|08003^Test Mode^99MRC||CBC+DIFF||||||F OBX|4|IS|01002^Ref Group^99MRC||||||||F OBX|5|NM|30525-0^Age^LN||PatientAge|yr|||||F OBX|6|ST|01001^Remark^99MRC||remark content....||||||F

Regards,

Arokia Swamy S

On Thu, Oct 17, 2024 at 8:24 AM Moses Mbadi @.***> wrote:

LIS Integration for: HumaCount 5D HC5D_LISInterfaceManual_V1.pdf https://github.com/PROTO-TYPE-SOLUTIONS/make-easy-hmis/files/14736429/HC5D_LISInterfaceManual_V1.pdf HumaLyte Plus 3 Humalyte-Plus3-LIS-InterfaceManual_V1.pdf https://github.com/PROTO-TYPE-SOLUTIONS/make-easy-hmis/files/14736455/Humalyte-Plus3-LIS-InterfaceManual_V1.pdf Maglumi 600 HumaStar 100

Hello! Now I'm trying to create LIS Integration with HumaCount 80ts analyzer. I've already implemented messages receiving, but I have problems with sending hl7 messages to analyzer, it does not reply. May be you have any documentation or advice to solve this problem? I would appreciate it.

@dayaDanya https://github.com/dayaDanya The HL7 message you are sending to the equipment should be correctly formatted. First, try to establish that the analyzer is communication with your LIS before sending the data. Then send the data and make sure it's formatted as explained in the guide attached. You can share your code I take a look.

— Reply to this email directly, view it on GitHub https://github.com/PROTO-TYPE-SOLUTIONS/easymed/issues/477#issuecomment-2418469208, or unsubscribe https://github.com/notifications/unsubscribe-auth/ACCRDLR7BCGGZHFH3XXHKVDZ343XVAVCNFSM6AAAAABJBYQCIGVHI2DSMVQWIX3LMV43OSLTON2WKQ3PNVWWK3TUHMZDIMJYGQ3DSMRQHA . You are receiving this because you commented.Message ID: @.***>