test

Latest version of an interface that's used to set up mice for high-precision proton irradiation based on registration of three necessary image sources:

Setup

If you do not yet have the Anaconda Python package manager, get it here. After installing, open an anaconda command prompt: screenshot_1

In the command prompt, create a new environment for RadiAIDD:

conda create -n RadiAIDD
conda activate  RadiAIDD

Install git and pip in this environment: conda install pip git

Now, create a new directory for RadiAIDD. Move into this directory and download + install this repository using

git clone https://github.com/jo-mueller/RadiAiDD.git
pip install -e .

Lastly, you can run RadiAIDD by simply typing radiaidd in the command line.

Necessary Input

As written above, you'll need these data input sources:

Workflow

The workflow is as follows.

  1. Define the beam isocenter. Navigate to the Radiography tab (top left) and click Define. In the opening IsoCenter definition window, load the image of the collimated beam spot and draw rectangle around it. It should look somehow like this: IsoCenter Note: You have to click the Lock-button to confirm the detection of the isocenter.

  2. Load Plan and treatment image: For this, head to the active positioning tab and load the respective images. You can change the grayscales by moving the mouse over the image, middle-mouse click and drag the mouse. When you are happy with the display, proceed to registration.

  3. Registration: Click Set default landmarks for both images and drag the appearing dots (you can change their sizes with the sliders below) to corresponding locations in both images. Click run Registration to display the resulting image transformation. The image overlay will show the same grayscale as the input images. The result should look somewhat like this: AcceptedRegistration Use the slider below the overlay images to flip back and forth between planing and treatment image.

  4. Target definition: In the target coordinates field, click the Get button to place the target marker in a random location in the plan image. If available, you can use the Show atlas button to display any provided overlays of the image. It may happen that, after loading the plan image, you will see only the atlas, and not the actual image. If this is the case, use the Flip Image button to exchange image and overlay. Drag&drop the target marker to your desired location in the planing image and click run Registration again to project the target coordinates into the treatment image. When you're good to go, click Accept Registration.

  5. Motor control: to be written

This gif shows the entire workflow: Workflow

Bug handling & known issues