Design files and instructions for manufacturing NP1.0, NP2.0alpha, and NP2.0 Neuropixels chronic recording enclosures.
Please reach out to us if you want to try this approach, we need your input. We are actively working on the instructions and trying to find vendors that can print holders reliably.
These probe enclosures use the dovetail to secure the probes, the design is at the current limit of the printing resolution of the SLA technology and for that reason, it is sensitive to the material, and to the washing and curing steps. After printing, the screw hole has to be tapped using a standard 1mm thread tapper.
The NP1.0 holder weights 1.2g, the NP2.0 holder weights 0.57g. All weights include the respective probe, chassis and cap, but no headstage.
There are 7 total steps to using this design for your experiments. Documentation is still being added but we will eventually cover them all here.
Here we describe how to tap printed parts, check the printing holders section for how to get the parts printed.
Sharpening probes makes that there is less resistance when driving probes during insertion and can prevent the need for duroctomies.
Soldering the ground and reference wires are an essential part of preparation to reduce noise. Neuropixels probes are less prone to noise than others because the digitization occurs at the probe. Here we discuss the optimal grounding strategies for using the multiple probes chronically. Some of this knowledge is guided by discussions with Bill Karsh (main developer of SpikeGLX) and Jan Putzeys (main developper of the acquisition system).
There are many ways of connecting multiple probes and ground to the animal that work, here we present one of them. The probes have 2 soldering points, the ground and the reference. Ideally we connect the ground from one of the probes to the animal (screw in the skull) to ground the animal. The reference from each probe can then be connected together and to a screw in the animal.
Technically, the best configuration separates the ground from the reference, we don't always do this and often connect the ground from one of the probes to the reference of all probes and attach that to a ground screw that is touching the dura.
When implanting many probes one can consider using the tip of the probe as reference. NP2.0 allow connecting the tip of 2 probes as reference (2 per headstage).When implanting less than 4 probes we often use the external reference as ground.
The design uses the "metal cap" (aka dovetail) to guide and secure the probe (through a screw in the rear of the holder). To use this implant design, we recommend you order the probes denoted "with metal cap" on the IMEC website. If you already have probes without the metal cap and want to use this design, you can have the metal dovetail machined yourself and then glue it to the probe. We will provide the design files for that part and instructions on how to glue it soon.
To mount probes, insert the probe with the shank first throught the dovetail, drive it all the way with the help of forceps.
Critical:
To "recover" do the oposite:
You may have to adjust the tolerances depending on the printing and washing protocol you use, we provide scripts - iLogic - to do this programatically from Autodesk Inventor. You can get an academic license for Inventor if you work in an academic institution at zero cost.
Do not implant probes without testing the tolerances. We tipically print an array of differenet tolerance values to find out the one that fits best, then use that tolerance for all holders.
We used a FormLabs Form 3+ printer and GreyPro resin to develop these protocols, please let us know if you use another printer or resin. Clear, and Rigid 10k, will NOT work (too brittle). Black V2 also worked in our hands but was not as good. Check #2 for an alternative material using a commercial vendor ( thanks @wangxuanyu0419 ).
Each version NP1.0, NP2.0 and NP2.0alpha has a folder in the repository. Print using the files in the stl_exports folder and using PreForm (we will provide files to print directly in the future; for now feel free to reachout to us for these files).
name | usage | vendor | cost estimate |
---|---|---|---|
Form 3+ | SLA printer | Formlabs | around 3k (2023) |
Form Wash | wash parts | Formlabs | around 650 (2023) |
Form Cure | cure parts | Formlabs | around 750 (2023) |
[Resin tank]() | 3d printer tank (needs to be replaced from time to time) | Formlabs | around 150 (2023) |
Grey Pro resin | resin to print the holders (needs to be replaced from time to time) | Formlabs | around 200 (2023) |
Isopropyl 99% | wash parts | multiple vendors | 150 per gallon |
Read the printer instructions for tricks on how to print, wash and cure. Washing times depend on how clean the isopropyl is and strongly affect the tolerances.
Tested on PreForm 3.32.0. Layer tickness 0.050mm with GreyPro resin.
Printing should take <7 h depending on how many parts are being printed. After printing, wash for 15-20 min depending on how clean is the isopropyl is, and let dry for at least 4 h (avoid sun light). If the parts feel "sticky" after this drying period, wash for 2 additional minutes. Cure the parts in Form Cure with the default settings for Grey Pro.
The following lab tools are required to assemble and implant a probe holder:
name | usage | vendor |
---|---|---|
Solder station | Solder wire to probes (if using external ground) | any |
Fine tipped forceps | Handling probes | any (example here) |
Small phillips screwdriver (M1) | Secure probe in holder | any |
Tap wrench | Tap the plastic holder | any |
M1 tap | Tap the plastic holder | any |
Hex key (todo, add size) | secure holder to stereotax | any |
Putty | Temporarily store assembled holders | todo: link |
Kopf surgical stereotax | stereotaxic surgery | Kopf (we will soon make stereotax arm attachments for other manufactures) |
Standard rodent surgical toolset | rodent surgery | any |
(Optional) Metal cap holder | Hold the probes for cleaning | IMEC (There are different versions for the 1.0 and 2.0 probes, get the correct one for the probes you have) |
(Optional) Thorlabs breadboard, posts and rotating post clamps | Hold probes for washing. (It can be nice to use Thorlabs parts to build a dedicated washing setup to keep surgical rigs unoccupied, but you can also just use your stereotax) | Thorlabs |
The following parts are also required to assemble and implant one probe, not including the 3D printed parts.
name | usage | vendor |
---|---|---|
Neuropixels probe (1.0/2.0alpha/2.0) with metal cap | record neural activity | IMEC |
M1x4 (NP2) or M1x5 (NP1) phillips head screw | secure probe to holder | any |
Nut (TODO, add size) | stereotax holder | any |
Bolt (TODO, add size) | stereotax holder | any |
(Optional) Kapton tape | secure probe and silver wire for soldering | TODO |
(Optional) Silver wire | grounding, if external grounding is desired | TODO |
These designs are not intended for commercial use or distribution.
Please reach out for feedback. Max Melin (mmelin@ucla.edu) and Joao Couto (couto@ucla.edu)